| Calibration
Sound
level meters should be
calibrated before and
after every use to
ensure that they are
working as they should.
The method used,
then, must be amenable
to field use.
One
way would be to expose
the meter to a known
sound pressure level, ie.,
to use an acoustical
calibrator. These are designed to fit closely over the microphone and can
be divided into two
groups:
Pistonphones
are precision
calibrators which should
be accurate to ±0.1
dB.
Their use is
generally confined to
laboratories.
Alternatively
a built-in or insert
voltage calibration can
be made.
This injects a
known, precise
electrical voltage into
the microphone circuit
which produces a known
response.
The drawback of
this approach is that it
would not detect any
change in the microphone
sensitivity.
There
are other, more precise,
laboratory methods of
calibration, but these
are used to calibrate
test laboratory
instruments rather than
normal measurement
instruments for use in
the field or in the
laboratory.
All
calibrators are used in
the same way.
The calibrator is
placed in position over
the microphone and the
SLM reading is noted; if
it is not the same as
the known output level
of the calibrator, then
adjustments are made to
the meter until it
displays the correct
reading.
In
older instruments, this
adjustment is usually
made using a screw in a
recess on the
instrument, the recessed
position preventing it
from being accidentally
knocked.
A large variation
(ie., more than 1 dB)
from the previous
calibration indicates a
fault in the instrument
which needs further
examination. Modern equipment incorporates electronic adjustment systems.
It
is important that the
calibration should
produce a sufficiently
high output as to be
unaffected by the
ambient noise level.
Background
noise may affect the
calibration.
It could either
produce an unsteady
reading on the SLM or
add to the indicated
level, with the result
that the meter would be
adjusted to a lower
level than is correct.
It is best to
choose as quiet a place
as possible for the
calibration, but in any
case the calibrator
signal must be at least
10 dB higher than the
ambient background
level.
The
meter should also
incorporate a ‘battery
check’ setting and a
‘low battery’
indicator, eg., a
flashing light or
digital display warning
of low battery levels.
Battery power is
consumed rapidly;
leaving the equipment
switched on overnight in
its case will usually
drain the batteries.
It is essential
to keep a spare set of
batteries with the
instrument.
In
addition to these
routine tests, the
instrument should
periodically be more
thoroughly tested in a
specialist laboratory.
Individual
equipment suppliers
should be consulted as
to how frequently they
recommend a full
laboratory calibration
of their equipment.
The HSE suggest
(in Noise Guide 3) that
the interval between
calibrations should not
normally be longer than
2 years).
»
Sound
Level Meters
»
Microphones
»
The
Dynamic Range
»
Calibration
»
Equipment
Standards
»
Using
SLM's in the Field
» Sound
Level Instruments
Stocked
»To add this site to your favourites
click here
|